KKK-A-1822 E
b.
Three see through inflatable face-masks, one each in adult, child, and infant sizes. Two
masks covering the full range of adult through pediatric sizes may be furnished in lieu of
three masks.
c.
Standard 15 mm/22 mm fittings.
d.
A true non-breathing valve.
e.
An oxygen reservoir system for delivery of 80-100 percent oxygen through an ancillary
oxygen inlet. The system shall be capable of accepting at least 15 liters/min. of oxygen
without valve jam.
3.12.2.2 OXYGEN POWERED, MANUALLY TRIGGERED (RESUSCITATOR) VALVE.
When specified (3.15.4 M20), an oxygen powered manually triggered (resuscitator) valve shall
be provided. This device shall have a see through inflatable face mask, interface with the action
wall outlet, via the specified quick disconnect device, and/or the dual port regulator/ flowmeter
of the portable oxygen system, and meet the following requirements:
a.
Provide oxygen for breathing and non-breathing patients.
b.
Provide manually triggered flow rates in accordance with the American Heart Association
(AHA), ASTM, or other recognized medical/standards organization(s).
c.
Provide not less than 85 percent oxygen.
d.
Be equipped with a safety valve having a release pressure setting of 60 cm of water
for adults.
3.12.2.3 AUTOMATIC RESUSCITATOR.
When specified (3.15.4 M29), an automatic resuscitator/volume ventilator shall be furnished.
The device's features and performance shall conform to AHA or other medical/standards
organization requirements. The device shall be furnished with a connection hose and
ambulance and/or portable oxygen system compatible quick disconnect fittings.
3.12.3
SUCTION ASPIRATOR, PRIMARY PATIENT.
Unless otherwise specified (see 6.2 nn), an electrically powered suction aspirator system
shall be furnished with an illuminated switch and a panel mounted, labeled, quick disconnect
inlet device on the EMT panel. The suction pump shall be located in an area that is
accessible but sound and vibration insulated from the patient compartment. The pump
shall be vented to the vehicle's exterior. When specified by the purchasers (see 6.2) for
gasoline powered vehicles, an engine powered suction system may be furnished and shall
include a hospital type self sealing plug-in outlet. The system shall include a corrosion
resistant reservoir having a volume of not less than 10.5L (640 cu. in.) which is connected
to the engine manifold through a check valve. All components, electrical, pressure and/or
vacuum, and other lines and accessories, shall be securely mounted yet readily accessible.
The electric type aspirator system shall be connected per Figure 5A or 5B. The aspirator
system shall provide a free air flow of at least 30 LPM and achieve a minimum of 300 mm
(11.81 in.) Hg vacuum within four seconds after the suction tube is closed. The rate of
flow and pressure of the engine vacuum aspirator system shall be attainable continuously
during the entire normal range of vehicle operation (i.e., accelerating, decelerating, at rest,
climbing hills, etc.). A vacuum control and a shut-off valve, or combination thereof, shall
be provided to adjust vacuum levels and to discontinue aspiration instantly. A vacuum
indicator gauge of 76 mm +/-13 mm (3" +/-0.5") in diameter, with numerical markers at
56